COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND
ARGUMENT 1 ARGUMENT 2
ARGUMENT 3 ARGUMENT 4 RESPONSE
4-20mA
Analog
Output
Settings
Sets/Reads Analog Outputs assignment,
4mA and 20mA Values. NOTE: 4mA and
20mA values have to be entered in
corresponding to Argument 2 type EU
(for Volum. in %FS). 0 - Disabled.
1-Volum. Flow: Current active
volumetric flow EU. 2-Mass Flow:
Current active volumetric flow EU.
3-Temperature: deg. C 4-Pressure: Pa.
22 AO <Value>
Channel: 1 or
2
<Value> Mode: [0-4]
0 - Disabled
1-Vol. Flow
2-Mass Flow
3-Temp.**
4-Pressure**
<Value>
4mA value
(in EU related
to Mode
value)
<Value>
20mA value
(in EU related
to Mode
value)
Ch1
AO1:0,0,100
Ch2
AO2:2,0.0,26.0
No Argument (Ret.
Current settings. For
Volum. Flow in %FS)
Ch1. AO1:1,0.0,1000.0
Ch2. AO2:2,0.0,26.0
Data Log Reads Data Log entries, Sets
Mode and Configuration of
the data logger. Argument 1 = C
DL Configuration (Argument2):
0 – Loop Enabled
1 – Loop Disabled
2 – Loop Disabled,
number of samples to record DL
Configuration (Argument3): Number of
samples: [1-15871] (has to be
submitted for all options, but only
applicable for Argument 2 value 2).
Argument 1 = V. Process Variable:
0-Volumetric Flow 1-Mass Flow
2-Temperature**. 3-Pressure**
Argument 1 = T
0 – Unconditional
1 – PV Alarm High Event
2 – PV Alarm Low Event
23 DL M
Mode:
E – DL Enable
D – DL Disable
<Value>
E or D
DLM:<value>
Example: DLM:E
No Argument (Ret.
Current settings)
DLM:<value>
Example: DLM:E
C
Configuration
<Value> [0-2]
<Value>
: [1-15871]
DLC:<value>,<value>
Example: DLC:0,1024
No Argument (Ret.
Current settings)
DLC:<value>,<value>
Example: DLC:0,1024
V
Process
Variable
<Value> [0-3]
DLV:<value> Example: DLV:0
No Argument (Ret.
Current settings)
DLV:<value>
Example: DLV:0
I
Sample
Interval in sec.
<Value> [1-3600]
DLI:<value> Example: DLI:1
(Ret. Current
settings)
DLI:<value> Example: DLI:1
T
DL Start/Stop
conditions
Thresholds
<Value>
Start Cond[0-2]
<Value>
Stop: [0-2]
DLT:<value>,<value>
Example: DLT:0,0
(Ret. Current
settings)
DLT:<value>,<value>
Example: DLT:0,0