Aalborg Vortex Insertion Manuale Utente Pagina 96

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 95
91
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND
ARGUMENT 1
ARGUMENT 2 ARGUMENT 3
ARGUMENT 4
RESPONSE
Real Time
Calendar
Read and set date/time
for battery backed
calendar.
16 C R
(read calendar) Returns
current date and time.
CR:MM/DD/YYYY,HH:MM:SS
Example:
CR:07/28/2010,12:11:52
D
(set date)
<Date string>
mm/dd/yyyy
CD:mm/dd/yyyy
Example: CD:07/28/2010
T
(set time)
<Time string>
hh:mm:ss
CT: hh:mm:ss
Example: CT:17:30:55
System
Events
Log
Read System Event
Log records from
EEPROM and sets
Events Log parameters
See list of the Diagnostic
Events below.
NOTE: Events higher
than 0xFFFF can not
be masked.
Read Log Index
command returns:
Date - MM/DD/YYYY
Time - HH:MM:SS
Event Code – [0-24]
17
SL
L
Loop (when max. index
exceeded rollover to
index 0)
(New Value) [E or D] SLL:<value> Example: SLL:D
No Argument
(Return current settings)
SLL:<value>
Example: SLL:D
S
Log Status: Current Index
[0-511] Rollover counter
SLS:<Index>,<Counter>
Example: SLS:97,0
M
System Event Log Mask
NOTE: Events higher than
0xFFFF can not be
masked.
<Value> 0x0000 To
0xFFFF (all 6 characters
are required)
SLM:<value>
Example: SLM:0x0013
No Argument
(Return current settings)
SLM:<value>
Example: SLM:0x0013
R
Read Log Index
<Value>
Index: [0-511]
SLR:Date,Time,Event
SLR:07/28/2010,12:11:52,22
C
Clear SysEvents Log
SLC
Flow Units
Selections
(defined in
#29)
Display/Change Meter
Flow Units: S - Standard
A - Actual
N - Normal
18
FC (New Value) [S,A,N]
FC: <Value> Example: FC:S
No Argument
(Returns current flow
units cond. settings)
FC: <Value>
Example:
FC:S
Vedere la pagina 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 122 123

Commenti su questo manuale

Nessun commento