Aalborg Vortex Insertion Manuale Utente Pagina 87

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 86
82
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND ARGUMENT 1 ARGUMENT 2
ARGUMENT 3
ARGUMENT 4 RESPONSE
Meter
Info
Requests meter
configuration info:
- full scale range
(L/min)
- Meter size (inches)
- T/P option support
(Y,N)
- 4-20 mA Output #1
configuration
- 4-20 mA Output #2
configuration
7 MI MI: 1000.0,0.75,Y,V,M
1000.0 – full scale L/min
0.75 – meter size [inch]
Y – T/P supported
N – T/P not supported
V – volumetric flow
M – mass flow
T – Temperature
P – Pressure
D – Density
O – Off (Disabled)
Flow
Alarms
Sets / reads the status
of the flow alarms.
Note: High and Low limits
for Volumetric flow
configuration have to be
entered in fraction %FS
notation (0.0 – 1.0). High
and Low limits for Mass
flow configuration have to
be entered in the currently
active mass flow units.
High alarm value has to
be more than Low alarm
value.
Alarm conditions:
Flow > High Limit = H
Flow < Low Limit = L
Low < Flow < High = N
8 FA C (flow configuration) V-volum. flow
M-mass flow
<Value>
(high limit)
<Value>
(low limit)
FAC:V,0.9,0.1
A (action delay in sec) <Value> [0-3600] FAA:<Value (sec)>
E (enable alarm) FA:E
D (disable alarm)* FA:D
R (read current
status)
FAR:N (no alarm)
FAR:H (high alarm)
FAR:L (low alarm)
S (Read current
settings)
FAS:M,C,H,L,D,B
where: M – mode (E/D)
C– Configuration (V/M)
H – High settings value
L – Low settings value
D – Action Delay (sec)
B – Latch mode (0-1)
L (Latch mode) <Value> (0-disabled*)
(1-enabl’d)
FAL:<Value> where:
Value = 0 – 1
Vedere la pagina 86